US EPA Method 8015 (modified): GC Analysis of Gasoline Range Organics (GRO) on VOCOL® after Purge & Trap using "M" Trap

suitable for GC

Faça loginpara ver os preços organizacionais e de contrato


About This Item

técnica(s)

gas chromatography (GC): suitable

parâmetros de teste

sample/matrix: Each analyte at 50 ppb in 5 mL water
purge trap: M trap (20079-U)
purge: 40 mL/min at 25 °C for 11 min
dry purge: 2 min
desorption preheat: 220 °C
desorption process: 40 mL/min at 225 °C for 2 min
bake: 260 °C for 10 min
transfer line/valve temp.: 110 °C
column: VOCOL, 30 m x 0.25 mm I.D., 1.5 μm (24205-U)
oven: 60 °C (2 min), 8 °C/min to 165 °C, 25 °C/min to 220 °C (5 min)
detector: FID, 250 °C
carrier gas: helium, 1.0 mL/min
injection: 30:1 split
sample: 0.75 mm I.D. SPME
inj. temp.: 200 °C

aplicação(ões)

environmental

Informações legais

VOCOL is a registered trademark of Merck KGaA, Darmstadt, Germany

Analito

Descrição

    Pentane
    2-Methylpentane
    2,2,4-Trimethylpentane
    Benzene
    Toluene
    Nonane
    Ethylbenzene
    m-Xylene
    p-Xylene
    o-Xylene
    Decane
    1,2,4-Trimethylbenzene
    Butylcyclohexane
    Naphthalene
    Ver tudo (14)

Escolha uma das versões mais recentes:

Certificados de análise (COA)

Lot/Batch Number

Lamentamos, não temos COA para este produto disponíveis online no momento.

Se precisar de ajuda, entre em contato Atendimento ao cliente

Já possui este produto?

Encontre a documentação dos produtos que você adquiriu recentemente na biblioteca de documentos.

Visite a Biblioteca de Documentos

Nossa equipe de cientistas tem experiência em todas as áreas de pesquisa, incluindo Life Sciences, ciência de materiais, síntese química, cromatografia, química analítica e muitas outras.

Entre em contato com a assistência técnica