0.5-3.0 L/min flow rate
32-113 °F temp. range (0-45 °C)
técnica(s)
active air sampling: suitable
energia da bateria
4.8 V, 14 hr life
aplicação(ões)
air monitoring
environmental
industrial hygiene
Categorias relacionadas
Descrição geral
Specifications | |
---|---|
Operating Flow Control | Volumetric flow rate held within ± 2.5% of set-point over the 1 to 3 L/min operating range (± 5% to 0.5 LPM), with automatic compenstation for battery voltage, altitude, temperature, and sample load changes. |
Flow Rate | Flow rate adjustable between 0.5 and 3.0 L/min; lower flow rates attainable (to 1 mL/min) with the optional Gemini Twin-Port Sampling accessory |
Flow Indication | Liquid crystal display (LCD) provides 0.01 L/min resolution |
Operating Range | 30 inches of water load upt to 2 L/min and 10 inches up to 3 L/min |
Flow Blockage Detection | Flow-fault LED comes on immedieatly when block is detected. Pump shuts down within 90 seconds if block is not cleared. |
Elapsed Time Readout | To 999 minutes in increments of 1 minute. Holds the last reading after either flow blockage or low battery shutdown and when pump is OFF or in HOLD mode. |
Operating Time | Varies with flow rate and sample device loading. |
Operating Temperature Limits | 32-113 °F (0-45 °C) |
Physical Weight | 23 oz. (652 g) with battery pack |
Dimensions | 2.25 in. D x 4.25 in. H x 4.0 in. W (5.7 cm D x 10.8 cm H x 10.2 cm) with battery pack |
Power Supply | 4.8 V battery pack of four NiMH cells (units mfg before October 2011 contain NiCd cells) |
Battery Pack Capacity | 2.8 amp hours (NiMH version); 2.4 amp hours (NiCD version) |
Battery Pack Recharge Time | 14-16 hours overnight with Omega Battery Charger |
Typical Battery Pack Life | 300 or more charging cycles |
Características e benefícios
- Patented electronic laminar flow control
- Internal secondary standard calibrates the pump continuously, maintaining OSHA compliance
- Check against a primary standard only once a month or every 200 hours for coal mine dust sampling
- Inlet filter traps moisture and particular matter
- Stainless steel-filled plastic case provides EMI/RFI protection
- Rated IP66 for resistance to water and dust ingress
- Exceptionally quiet operation
- Flow-Fault Indicator shows any sample inlet blockage
- Run/Hold feature allows "time out" during sampling process
- Elapsed time readout and low-battery warning
- Low flow and dual tube sampling available with Gemini Twin-Port Sampler (28118-U)
Adequação
Approvals
- NIOSH certified for coal mine dust sampling (TC-74-030)
- MSHA certified as intrinsically safe for underground use (Approval # 2G-3924-1).
- UL approved as intrinsically safe for use in hazardous locations, Class I, Groups A, B, C, D; Class II, Groups E, F and G; Class III, Division I locations.
Informações legais
necessário, mas não fornecido
produto relacionado
Nº do produto
Descrição
Preços
recomendado
Nº do produto
Descrição
Preços
Código de classe de armazenamento
11 - Combustible Solids
Classe de risco de água (WGK)
WGK 3
Ponto de fulgor (°F)
Not applicable
Ponto de fulgor (°C)
Not applicable
Escolha uma das versões mais recentes:
Certificados de análise (COA)
Lamentamos, não temos COA para este produto disponíveis online no momento.
Se precisar de ajuda, entre em contato Atendimento ao cliente
Já possui este produto?
Encontre a documentação dos produtos que você adquiriu recentemente na biblioteca de documentos.
Nossa equipe de cientistas tem experiência em todas as áreas de pesquisa, incluindo Life Sciences, ciência de materiais, síntese química, cromatografia, química analítica e muitas outras.
Entre em contato com a assistência técnica