HPLC Analysis of Benzodiazepines on Ascentis® Express Phenyl-Hexyl with Acetonitrile Mobile Phase

HPLC Analysis of Benzodiazepines on Ascentis® Express Phenyl-Hexyl with Acetonitrile Mobile Phase application for HPLC

材料

分析カラム

製品番号
詳細
価格

Ascentis® Express フェニルヘキシル、2.7 μm HPLCカラム

2.7 μm particle size, L × I.D. 5 cm × 4.6 mm

標準

製品番号
詳細
価格

デスメチルジアゼパム

analytical standard

オキサゼパム 溶液

1 mg/mL in methanol, ampule of 1 mL, certified reference material, Cerilliant®

アルプラゾラム 溶液

1.0 mg/mL in methanol, ampule of 1 mL, certified reference material, Cerilliant®

クロナゼパム 溶液

1.0 mg/mL in methanol, ampule of 1 mL, certified reference material, Cerilliant®

ジアゼパム 溶液

1.0 mg/mL in methanol, ampule of 1 mL, certified reference material, Cerilliant®

CONDITIONS

column

Ascentis Express Phenyl-Hexyl, 5 cm x 4.6 mm I.D., 2.7 μm particles (53348-U)

mobile phase

[A] water; [B] acetonitrile; (62:38, A:B)

flow rate

1.4 mL/min

pressure

2600 psi (179 bar)

column temp.

35 °C

detector

UV, 250 nm

injection

2 μL

sample

0.1 g/mL of each in 20% methanol

詳細

アナリシスノート

The Phenyl-Hexyl phase of the Ascentis Express product line displays the best selectivity of this sample; selectivity is markedly affected by choice of organic modifier. The application also demonstrates feasibility of fast LC with Ascentis Express by virtue of low backpressure and high efficiency.