コンテンツへスキップ
Merck

HPLC Analysis of Caffeine and Taurine on Ascentis® Express HILIC Column using Evaporative Light Scattering Detection (ELSD)

HPLC Analysis of Caffeine and Taurine on Ascentis® Express HILIC Column using Evaporative Light Scattering Detection (ELSD) application for HPLC

材料

Standard

製品番号
詳細
価格

タウリン

≥99%

カフェイン

certified reference material, TraceCERT®, Manufactured by: Sigma-Aldrich Production GmbH, Switzerland

分析カラム

製品番号
詳細
価格

Ascentis® Express HILIC、2.7 μm HPLCカラム

2.7 μm particle size, L × I.D. 5 cm × 3 mm

CONDITIONS

column

Ascentis Express HILlC, 5 cm x 3.0 mm I.D., 2.7 μm particles (53967-U)

mobile phase

[A] 100 mM ammonium formate, pH 3.0 with formic acid; [B] acetonitrile; (10:90, A:B)

flow rate

0.6 mL/min

column temp.

25 °C

detector

ELSD, 55 °C, 3.5 bar nitrogen

injection

2 μL

sample

1000 μg/mL in 10:90, water: acetonitrile

詳細

アナリシスノート

This application demonstrates the suitability of Ascentis Express HILIC for the efficient separation of caffeine and taurine. Elution is by aqueous normal phase (ANP) as an increase in organic will increase retention.

法的情報