GC Analysis of Water (100% Standard) on Watercol 1460

GC Analysis of Water (100% Standard) on Watercol™ 1460 suitable for GC

CONDITIONS

column

Watercol 1460, 30 m x 0.25 mm I.D., 0.20 μm (29695-U)

oven

60 °C (12 min)

inj. temp.

250 °C

detector

TCD, 200 °C

carrier gas

helium, 26 cm/sec

injection

0.1 μL, 100:1 split

liner

4 mm I.D., split type, cup design

sample

water

Opis

Komentarz do analizy

The "Watercol" series of ionic liquid capillary GC columns are characterized by their ability to produce a sharp peak shape for water, sharp enough so that water can be integrated and quantified. This application shows a calibration standard (100% water) on Watercol 1460.