Przejdź do zawartości
Merck

HPLC Analysis of PAHs in Water on SUPELCOSIL LC-PAH after SPME using 100 μm PDMS Fiber

HPLC Analysis of PAHs in Water on SUPELCOSIL™ LC-PAH after SPME using 100 μm PDMS Fiber application for SPME, application for HPLC

Materiały

Włókno SPME

Numer produktu
Opis
Cennik

SPME fiber assembly Polydimethylsiloxane (PDMS)

df 100 μm(PDMS, for use with autosampler, needle size 24 ga

kolumna analityczna

Numer produktu
Opis
Cennik

SUPELCOSIL LC-PAH HPLC Column

5 μm particle size, L × I.D. 15 cm × 4.6 mm

CONDITIONS

sample/matrix

5 μL PAH mix (Cat. No. 4-8743) in 5 mL water

SPME fiber

polydimethylsiloxane, 100 μm (57301)

extraction

immersion, 30 min (rapid stirring)

desorption process

static, 200 μL acetonitrile:water, 40:60, 2 min

column

SUPELCOSIL LC-PAH, 15 cm × 4.6 mm I.D., 5 μm particles (58318)

gradient

0 min - 50% A; 5 min - 50% A; 30 min - 100% A; 45 min - 100% A

flow rate

0-2 min: 0.2 mL/min 2-45 min: 1.0 mL/min

mobile phase

[A] acetonitrile:[B] water

detector

UV, 254 nm

Opis

Informacje prawne

SUPELCOSIL is a trademark of Sigma-Aldrich Co. LLC