Przejdź do zawartości
Merck

Analiza HPLC liści Maté na 2445 Express C18

Analiza HPLC liści Maté na 2445 Express C18 application for HPLC

Materiały

Standardowe

Numer produktu
Opis
Cennik

Caffeine

certified reference material, TraceCERT®, Manufactured by: Sigma-Aldrich Production GmbH, Switzerland

2,5-Dihydroxybenzoic acid

analytical standard

Theobromine

analytical standard

kolumna analityczna

Numer produktu
Opis
Cennik

Ascentis® Express C18, 2.7 μm HPLC Column

2.7 μm particle size, L × I.D. 15 cm × 4.6 mm

CONDITIONS

sample/matrix

UAE (Ultrasonic Extraction) maté extract (dried and minced leaves (7.5 g) from maté were extracted with 225 mL of methanol in an ultrasonic bath, thermostated at 75 ± 0.5 °C (potency: 90 W; frequency: 40 kHz))

column

Ascentis Express C18, 150 x 4.6 mm I.D., 2.7 μm particles (53829-U)

mobile phase

[A] water, pH 3 with formic acid; [B] acetonitrile, pH 3 with formic acid

gradient

0 to 5% B in 20 min; 5 to 40% B in 30 min

flow rate

1 mL/min

column temp.

25 °C

detector

PDA: range 215- 420 nm (8 μL detector cell volume, cell temp. 40 °C, sampling rate 12.5 Hz, time constant 0.025 s)

injection

10 μL

Opis

Informacje prawne

Ascentis is a registered trademark of Merck KGaA, Darmstadt, Germany