HPLC Analysis of Acrylamide in Fried Potato Chips on Discovery® HS F5 after SPE using Discovery® DSC-MCAX and DSC-18

HPLC Analysis of Acrylamide in Fried Potato Chips on Discovery® HS F5 after SPE using Discovery® DSC-MCAX and DSC-18 application for SPE, application for HPLC

CONDITIONS

sample preparation

SPE (Solid Phase Extraction)

sample/matrix

Finely grind 2.0 g of potato chips, place in 20 mL vial and add 10 mL water. Mix using vortex until mixture forms a thick paste. (Place sample in centrifuge tubes and centrifuge at 16.1 rcf for 5 min. Extract aqueous portion of sample leaving oil layer and solids.)

SPE tube/cartridge

Discovery DSC-MCAX, 300 mg/3 mL (52784-U)

SPE tube/cartridge

Discovery DSC-18, 1g/6mL (52606-U)

condition

connect a Discovery DSC-MCAX SPE tube to the top of a Discovery DSC-18 SPE tube using an SPE tube adapter, pass 1 mL methanol, then 1 mL DI water through both SPE tubes until the tubes are dry

sample addition

1 mL of aqueous extract

washing

1 mL DI water

elution

removing upper MCAX SPE tube, pass 2 mL methanol through lower DSC-18 SPE tube

eluate post-treatment

evaporate to dryness with nitrogen at 30 °C, reconstitute in 0.5 mL DI water

column

Discovery HS F5, 15 cm x 4.6 mm I.D., 3 μm particles (567507-U)

mobile phase

100% ultra-pure water

column temp.

35 °C

detector

ESI+

injection

5 μL