Przejdź do zawartości
Merck

ASTM D5501: GC Analysis of Ethanol on Petrocol® DH 150

ASTM D5501: GC Analysis of Ethanol on Petrocol® DH 150 suitable for GC

Materiały

kolumna analityczna

Numer produktu
Opis
Cennik

Petrocol® DH 150 Capillary GC Column

L × I.D. 150 m × 0.25 mm, df 1.00 μm

CONDITIONS

column

Petrocol DH 150, 150 m x 0.25 mm I.D., 1.0 μm (24155)

oven

60 °C (15 min), 30 °C/min to 250 °C (23 min)

inj. temp.

300 °C

detector

FID, 250 °C

carrier gas

helium, 24 cm/sec @ 60 °C

injection

0.5 μL, 150:1 split

liner

4 mm I.D., single taper

sample

Denatured Fuel Ethanol Solution 4, from ASTM D5501 Denatured Fuel Ethanol Standards Kit (40361-U)

Opis

Informacje prawne

Petrocol is a registered trademark of Merck KGaA, Darmstadt, Germany