84-109
HPLC Analysis of Triazine Herbicides in Grass on SUPELCOSIL™ LC-8-DB after SPE using Supelclean™ LC-SCX
application for SPE, application for HPLC
About This Item
metody
HPLC: suitable
solid phase extraction (SPE): suitable
parametry testowe
sample preparation: SPE (Solid Phase Extraction)
sample/matrix: fresh grass clippings spiked with herbicides at 2 ppm
SPE tube/cartridge: Supelclean LC-SCX, 500 mg/3 mL (57018)
condition: 1 mL methylene chloride
sample addition: 2 mL grass extract. Wash with 2 x 2 mL acetonitrile.
drying: Dry packing for 5 min. under nitrogen purge
washing: Dry packing for 5 min under nitrogen purge; wash with 2 x 2 mL DI water
elution: 1.5 mL methanol
eluate post-treatment: dilute to 2 mL with DI water
column: SUPELCOSIL LC-8-DB, 15 cm × 4.6 mm I.D., 5 μm (58347-U) with guard column (59553)
mobile phase: [A] acetonitrile; [B] water; (45:55, A:B)
flow rate: 1.5 mL/min
detector: UV, 254 nm
injection: 100 μL
przydatność
application for HPLC
application for SPE
Zastosowanie
environmental
Zastosowanie
5 g fresh grass clippings spiked with 2 ppm each herbicide. Add 4 g anhydrous sodium sulfate and 20 mL methylene chloride:acetone (80:20). Shake 20 min and allow mixture to stand 1 min.
Informacje prawne
Analit
kolumna analityczna
prekolumna
probówka lub płytka SPE
Wybierz jedną z najnowszych wersji:
Certyfikaty analizy (CoA)
Przepraszamy, ale COA dla tego produktu nie jest aktualnie dostępny online.
Proszę o kontakt, jeśli potrzebna jest pomoc Obsługa Klienta
Masz już ten produkt?
Dokumenty związane z niedawno zakupionymi produktami zostały zamieszczone w Bibliotece dokumentów.
Nasz zespół naukowców ma doświadczenie we wszystkich obszarach badań, w tym w naukach przyrodniczych, materiałoznawstwie, syntezie chemicznej, chromatografii, analityce i wielu innych dziedzinach.
Skontaktuj się z zespołem ds. pomocy technicznej