G003784
Volatile Fraction of Roasted Coffee Beans Using SPME on Omegawax 250
application for SPME, suitable for GC
About This Item
tecniche
gas chromatography (GC): suitable
solid phase microextraction (SPME): suitable
Parametri analitici
sample/matrix: 2 g of coffee beans
SPME fiber: 50/30 μm divinylbenzene/Carboxen on polydimethylsiloxane on a StableFlex fiber (57329-U)
extraction: headspace, with a pre-equilibration period of 10 min. at 60 °C, and then a fiber exposure of 40 min. at 60 °C, agitated in an alternate clockwise-anticlockwise rotation mode at 500 rpm
desorption process: 260 °C for 5 min
column: Omegawax 250, 30 m x 0.25 mm I.D., 0.25 μm (24136)
oven: 40 °C (5 min), 4 °C/min to 230 °C, 50 °C/min to 280 °C (2 min)
inj. temp.: 260 °C
detector: MS, m/z 40-400
carrier gas: helium, 34.8 cm/sec constant
injection: splitless
scan range: m/z 40-400
sample: Roasted coffee
liner: 0.75 mm I.D. SPME/Splitless (2633901)
Compatibilità
application for SPME
applicazioni
food and beverages
Risultati analitici
Note legali
Analita
Colonna analitica
Componente della fase mobile
Fibra per SPME
Prodotti correlati
Standard
Scegli una delle versioni più recenti:
Certificati d'analisi (COA)
Ci dispiace, ma al momento non ci sono COA disponibili online per questo prodotto.
Se ti serve aiuto, non esitare a contattarci Servizio Clienti
Possiedi già questo prodotto?
I documenti relativi ai prodotti acquistati recentemente sono disponibili nell’Archivio dei documenti.
Il team dei nostri ricercatori vanta grande esperienza in tutte le aree della ricerca quali Life Science, scienza dei materiali, sintesi chimica, cromatografia, discipline analitiche, ecc..
Contatta l'Assistenza Tecnica.