Passa al contenuto
Merck
Tutte le immagini(1)

Key Documents

G005229

US EPA Method 8015 (modified): GC Analysis of Gasoline Range Organics (GRO) on VOCOL® after Purge & Trap using "M" Trap

suitable for GC

Autenticatiper visualizzare i prezzi riservati alla tua organizzazione & contrattuali


About This Item

tecniche

gas chromatography (GC): suitable

Parametri analitici

sample/matrix: Each analyte at 50 ppb in 5 mL water
purge trap: M trap (20079-U)
purge: 40 mL/min at 25 °C for 11 min
dry purge: 2 min
desorption preheat: 220 °C
desorption process: 40 mL/min at 225 °C for 2 min
bake: 260 °C for 10 min
transfer line/valve temp.: 110 °C
column: VOCOL, 30 m x 0.25 mm I.D., 1.5 μm (24205-U)
oven: 60 °C (2 min), 8 °C/min to 165 °C, 25 °C/min to 220 °C (5 min)
detector: FID, 250 °C
carrier gas: helium, 1.0 mL/min
injection: 30:1 split
sample: 0.75 mm I.D. SPME
inj. temp.: 200 °C

applicazioni

environmental

Note legali

VOCOL is a registered trademark of Merck KGaA, Darmstadt, Germany

Analita

Descrizione

    Pentane
    2-Methylpentane
    2,2,4-Trimethylpentane
    Benzene
    Toluene
    Nonane
    Ethylbenzene
    m-Xylene
    p-Xylene
    o-Xylene
    Decane
    1,2,4-Trimethylbenzene
    Butylcyclohexane
    Naphthalene
    Vedi tutto (14)

Standard


Scegli una delle versioni più recenti:

Certificati d'analisi (COA)

Lot/Batch Number

Ci dispiace, ma al momento non ci sono COA disponibili online per questo prodotto.

Se ti serve aiuto, non esitare a contattarci Servizio Clienti

Possiedi già questo prodotto?

I documenti relativi ai prodotti acquistati recentemente sono disponibili nell’Archivio dei documenti.

Visita l’Archivio dei documenti

Il team dei nostri ricercatori vanta grande esperienza in tutte le aree della ricerca quali Life Science, scienza dei materiali, sintesi chimica, cromatografia, discipline analitiche, ecc..

Contatta l'Assistenza Tecnica.