CONDITIONS
column
SPB-608, 30 m x 0.25 mm I.D., 0.25 μm (24103-U)
oven
80 °C (2 min), 15 °C/min to 250 °C, 8 °C/min to 300 °C (35 min)
inj. temp.
300 °C
MSD interface
300 °C
scan range
m/z 40 - 450
carrier gas
helium, 1.2 mL/min. constant
injection
0.5 μL, splitless (1 min)
sample
each analyte at 10 μg/mL in methylene chloride