Skip to Content
MilliporeSigma

HPLC Analysis of Chlorpheniramine Enantiomers on Astec® CYCLOBOND I 2000

HPLC Analysis of Chlorpheniramine Enantiomers on Astec® CYCLOBOND I 2000 application for HPLC

Materials

Analytical column

Product No.
Description
Pricing

Astec® CYCLOBOND I 2000 Chiral HPLC Column

5 μm particle size, L × I.D. 25 cm × 4.6 mm

Standard

Product No.
Description
Pricing

S-(+)-Chlorpheniramine maleate salt

(±)-Chlorpheniramine maleate salt

≥99% (perchloric acid titration)

CONDITIONS

column

CYCLOBOND I 2000, 25 cm x 4.6 mm I.D., 5 μm particles (20024AST)

mobile phase

[A] 1% triethylamine acetate, pH 4.1; [B] acetonitrile; (90:10, A:B)

flow rate

1 mL/min

column temp.

23 °C

detector

UV, 254 nm

injection

2 μL

sample

5 mg/mL in methanol

Description

Legal Information

Astec is a registered trademark of Merck KGaA, Darmstadt, Germany
null