HPLC Analysis of an Energy Drink on Ascentis® Express HILIC (UV Detection)

HPLC Analysis of an Energy Drink on Ascentis® Express HILIC (UV Detection) application for HPLC

Materials

analytical column

Product No.
Description
Pricing

Ascentis® Express HILIC, 2.7 μm HPLC Column

2.7 μm particle size, L × I.D. 10 cm × 3 mm

standard

Product No.
Description
Pricing

Caffeine

Sigma Reference Standard, vial of 250 mg

Sorbic acid

99.0-101.0% anhydrous basis

Pyridoxine Hydrochloride (B6)

analytical standard

Nicotinamide (Niacinamide)

analytical standard

Riboflavin (B2)

analytical standard

SS Benzoic Acid

2000 μg/mL in dichloromethane, analytical standard

CONDITIONS

column

Ascentis Express HILlC, 10 cm x 3.0 mm I.D., 2.7 μm particles (53970-U)

mobile phase

[A] 100 mM ammonium acetate, pH 5.0 with acetic acid; [B] water; [C] acetonitrile; (09:01:90, A:B:C)

flow rate

0.6 mL/min

pressure

812 psi (56 bar)

column temp.

35 °C

detector

UV, 254 nm

injection

2 μL

sample

Dilute 1:9 in acetonitrile

Description

Analysis Note

This application demonstrates the suitability of Ascentis Express HILIC for the efficient separation of common drink additives used in caffeinated energy drinks. The analysis was performed with UVand ELSD detectors in series, this chromatogram being from the UV.