GC Analysis of Oxygen, Nitrogen, and Carbon Dioxide on Carboxen®-1010 PLOT

GC Analysis of Oxygen, Nitrogen, and Carbon Dioxide on Carboxen®-1010 PLOT suitable for GC

Materials

Analytical column

Product No.
Description
Pricing

Carboxen®-1010 PLOT Capillary GC Column

L × I.D. 30 m × 0.53 mm, average thickness 30 μm

CONDITIONS

column

Carboxen-1010 PLOT, 30 m x 0.53 mm I.D. (25467)

oven

35 °C (7.0 min), 24 °C/min to 225 °C

inj. temp.

200 °C

detector

TCD, 230 °C, Methanizer, 375 °C, and FID, 230 °C in series

carrier gas

helium, 3.0 mL/min

injection

10 μL, direct using a sample valve, 150 °C

sample

Oxygen at 20% and carbon dioxide at 1.0% in nitrogen