Skip to Content
MilliporeSigma

HPLC Analysis of DNA on Proteomix® WAX-NP5

HPLC Analysis of DNA on Proteomix® WAX-NP5 application for HPLC

Materials

analytical column

Product No.
Description
Pricing

Proteomix® WAX-NP5 Weak Anion Exchange Column

L × I.D. 15 cm × 4.6 mm, 5 μm particle size

CONDITIONS

column

Proteomix WAX-NP5, 15 cm x 4.6 mm I.D., 5 μm particles (Z777262)

mobile phase

(A) 20 mmol Tris, pH 6.0; (B) A + 1.0 M NaCl

gradient

20 to 100% B in 40 min

flow rate

0.6 mL/min

pressure

1015 psi (70 bar)

temp.

ambient

detector

UV, 260 nm

injection

5 μL

sample

DNA

Description

Legal Information