GC Analysis of Closely Eluting Analytes from US EPA Method 8270 on SLB®-5ms (15 m x 0.10 mm I.D., 0.10 μm)

GC Analysis of Closely Eluting Analytes from US EPA Method 8270 on SLB®-5ms (15 m x 0.10 mm I.D., 0.10 μm) suitable for GC

Materials

Analytical column

Product No.
Description
Pricing

SLB®-5ms Capillary GC Column

L × I.D. 15 m × 0.10 mm, df 0.10 μm

CONDITIONS

column

SLB-5ms, 15 m x 0.10 mm I.D., 0.10 μm (28466-U)

oven

40 °C (2 min), 22 °C/min to 240 °C, 10 °C/min to 330 °C (1 min)

inj. temp.

250 °C

detector

FID, 330 °C

carrier gas

helium, 30 cm/sec @ 200 °C, set using methane

injection

0.1 μL splitless (0.75 min)

liner

2 mm I.D., straight

sample

50 ng on-column of a 72 component semivolatile standard and 8 surrogate compounds, plus 6 internal standards (at 40 ng on-column), in methylene chloride