Przejdź do zawartości
Merck
Wszystkie zdjęcia(1)

Key Documents

G005610

HPLC Analysis of Withania Extract on Ascentis® Express Phenyl-Hexyl

application for HPLC

Zaloguj sięWyświetlanie cen organizacyjnych i kontraktowych


About This Item

metody

HPLC: suitable

parametry testowe

column: Ascentis Express Phenyl-Hexyl, 10 cm x 2.1 mm I.D., 2.7 μm particles (53336-U)
mobile phase: [A] water; [B] acetonitrile
gradient: 20 to 100% B in 10 min; held at 100% B for 0.5 min
flow rate: 0.3 mL/min
pressure: 160 bar (2320 psi)
column temp.: 35 °C
detector: UV, 227 nm
injection: 5 μL
sample: 20 mg/mL in water:acetonitrile (50:50); sonicate 15 minutes; filter 0.45 μm;
dilute to water:acetonitrile (80:20)

przydatność

application for HPLC

Zastosowanie

vitamins, nutraceuticals, and natural products

Komentarz do analizy

This application demonstrates the suitability of Ascentis Express Phenyl-Hexyl for the efficient separation of withania extract.

Informacje prawne

Ascentis is a registered trademark of Merck KGaA, Darmstadt, Germany
This page may contain text that has been machine translated.

Analit

Opis

    Withanoside IV
    Withanoside V
    Withaferin A
    12-Deoxywithastramonolide
    Withanolide A
    Withanone
    Withanolide B

Wybierz jedną z najnowszych wersji:

Certyfikaty analizy (CoA)

Lot/Batch Number

Przepraszamy, ale COA dla tego produktu nie jest aktualnie dostępny online.

Proszę o kontakt, jeśli potrzebna jest pomoc Obsługa Klienta

Masz już ten produkt?

Dokumenty związane z niedawno zakupionymi produktami zostały zamieszczone w Bibliotece dokumentów.

Odwiedź Bibliotekę dokumentów

Nasz zespół naukowców ma doświadczenie we wszystkich obszarach badań, w tym w naukach przyrodniczych, materiałoznawstwie, syntezie chemicznej, chromatografii, analityce i wielu innych dziedzinach.

Skontaktuj się z zespołem ds. pomocy technicznej