Przejdź do zawartości
Merck
Wszystkie zdjęcia(1)

Key Documents

G004366

HPLC Analysis of Verapamil and Metabolites in Rat Plasma on Ascentis® Express C18 after Phospholipid Removal using HybridSPE®-Phospholipid

application for SPE, application for LC-MS

Zaloguj sięWyświetlanie cen organizacyjnych i kontraktowych


About This Item

metody

LC/MS: suitable
solid phase extraction (SPE): suitable

parametry testowe

SPE well plate: HybridSPE-Precipitation 96-well Plate, 50 mg/well (575656-U)
sample preparation: SPE (Solid Phase Extraction)
column: Ascentis Express C18, 5 cm x 2.1 mm I.D., 2.7 μm particles (53822-U)
sample/matrix: Rat plasma spiked with verapamil and metabolites at 10 ng/mL
mobile phase: [A] water; [B] acetonitrile, pH 2.55 adjust with formic acid (30:70, A:B)
flow rate: 0.6 mL/min
column temp.: 35 °C
detector: TOF/MS
injection: 1 μL
sample addition: 100 μL spiked rat plasma followed by 300 μL 1% formic acid in acetonitrile
elution: apply vacuum

przydatność

application for LC-MS
application for SPE

Zastosowanie

clinical

Informacje prawne

Ascentis is a registered trademark of Merck KGaA, Darmstadt, Germany
HybridSPE is a registered trademark of Merck KGaA, Darmstadt, Germany
This page may contain text that has been machine translated.

probówka lub płytka SPE


Wybierz jedną z najnowszych wersji:

Certyfikaty analizy (CoA)

Lot/Batch Number

Przepraszamy, ale COA dla tego produktu nie jest aktualnie dostępny online.

Proszę o kontakt, jeśli potrzebna jest pomoc Obsługa Klienta

Masz już ten produkt?

Dokumenty związane z niedawno zakupionymi produktami zostały zamieszczone w Bibliotece dokumentów.

Odwiedź Bibliotekę dokumentów

Nasz zespół naukowców ma doświadczenie we wszystkich obszarach badań, w tym w naukach przyrodniczych, materiałoznawstwie, syntezie chemicznej, chromatografii, analityce i wielu innych dziedzinach.

Skontaktuj się z zespołem ds. pomocy technicznej