796-0150
HPLC Analysis of Acidic Herbicides in Water on a Polymeric C18 Column after SPE using Supelclean™ ENVI™-Carb
application for SPE, application for HPLC
About This Item
Polecane produkty
metody
HPLC: suitable
solid phase extraction (SPE): suitable
parametry testowe
sample preparation: SPE (Solid Phase Extraction)
sample/matrix: Fresh 1 L water samples, dechlorinated with sodium thiosulfate when necessary, at ambient temperature and pH.
SPE tube/cartridge: Supelclean ENVI-Carb, 250 mg/6 mL (57092)
condition: 10 mL DI water at 20 mL/min
sample addition: 0.9 L water sample at 20 mL/min
drying: 10 min using clean nitrogen
washing: 10 mL DI water at 20 mL/min
elution: 10 mL 0.1% phosphoric acid in methylene chloride:acetonitrile (80:20) at 5 mL/min
column: polymeric-coated silica-based PAH specialty column, 20 cm × 3 mm I.D., 5 μm (Supelco equivalent, SUPELCOSIL LC-PAH, available upon request)
mobile phase: gradient, [A]: 0.05% phosphoric acid in DI water; [B]: acetonitrile
flow rate: 0.5 mL/min
column temp.: 50 °C
detector: photodiode array- peak width: 0.053 min, sampling interval: 0.320 sec, monitor 210 nm & 225 nm
injection: 10 μL of extract (4-5 ppb each analyte in water)
przydatność
application for HPLC
application for SPE
Zastosowanie
environmental
Zastosowanie
Informacje prawne
Analit
kolumna analityczna
probówka lub płytka SPE
Wybierz jedną z najnowszych wersji:
Certyfikaty analizy (CoA)
Przepraszamy, ale COA dla tego produktu nie jest aktualnie dostępny online.
Proszę o kontakt, jeśli potrzebna jest pomoc Obsługa Klienta
Masz już ten produkt?
Dokumenty związane z niedawno zakupionymi produktami zostały zamieszczone w Bibliotece dokumentów.
Nasz zespół naukowców ma doświadczenie we wszystkich obszarach badań, w tym w naukach przyrodniczych, materiałoznawstwie, syntezie chemicznej, chromatografii, analityce i wielu innych dziedzinach.
Skontaktuj się z zespołem ds. pomocy technicznej