추천 제품
ligand
alkali-stabilized protein A-derived (E. coli)
포장
pack of 1 × 5 mL
제조업체/상표
Cytiva 11-0034-94
저장 조건
(20% Ehtanol)
(20% Ethanol)
파라미터
<20 mL/min flow rate
42 psi
베드 크기
16 mm × 25 mm
베드 용적
5 mL
컬럼 I.D.
16 mm
Matrix
highly cross-linked agarose
평균 부품 크기
85 μm (d₅₀v is the median particle size of the cumulative volume distribution)
cleaning
2-13.7
작동 범위
3-12
용량
≥30 mg binding capacity (humanlgG/ml at 2.4 min residence time)
적합성
suitable for bioprocess medium
저장 온도
2-8°C
유사한 제품을 찾으십니까? 방문 제품 비교 안내
관련 카테고리
일반 설명
Determined at 10% breakthrough by frontal analysis at a mobile phase velocity of 500 cm/h in a column with a bed height of 20 cm.
HiTrap® MabSelect™ SuRe™ columns are prepacked with MabSelect™ SuRe, a BioProcess medium for purification of monoclonal antibodies from large sample volumes. The novel alkali-tolerant recombinant Protein A ligand allow cleaning-in-place (CIP).
HiTrap® MabSelect™ columns are prepacked with MabSelect™ a bioprocess medium for capuring of Mabs from large sample volumes.
HiTrap® MabSelect™ columns are prepacked with MabSelect™ a bioprocess medium for capuring of Mabs from large sample volumes.
pH below 3 is sometimes required to elute strongly bound IgG species. However, protein ligands may hydrolyze at very low pH.
특징 및 장점
- Withstands rigorous CIP and sanitization procedures with 0.1 to 0.5 M NaOH.
- Novel ligand design gives enhanced protease resistance resulting in lower ligand leakage.
- Oriented coupling of recombinant Protein A to the matrix via an engineered C-terminal cysteine enhances IgG binding capacity.
- Fewer regulatory concerns due to the total absence of mammalian culture in the ligand production and purification.
- Prepacked HiTrap® columns for process development, screening of purification conditions, and small-scale purification of Mabs.
분석 메모
To view the Certificate of Analysis for this product, please visit www.cytiva.com.
기타 정보
1)The pressure over the packed bed varies depending on a range of parameters such as the characteristics of the chromatography medium and the column tubing used.
The pressure over the packed bed varies depending on a range of parameters such as the characteristics of the chromatography medium and the column tubing used.
법적 정보
Free MabSelect SuRe Ligand
"MabSelect SuRe Ligand Restricted License" and "Cys-rProtein A Ligand Restricted License" are protected by the following patents and equivalent patents and patent applications in other countries: US 6,399,750 and EP 1123389. A free, non-transferable limited license to use this product for internal analytical purposes only accompanies the purchase of the product from a Cytiva company and its licensed distributors. Any other use will require a separate license from a Cytiva company.
"MabSelect SuRe Ligand Restricted License" and "Cys-rProtein A Ligand Restricted License" are protected by the following patents and equivalent patents and patent applications in other countries: US 6,399,750 and EP 1123389. A free, non-transferable limited license to use this product for internal analytical purposes only accompanies the purchase of the product from a Cytiva company and its licensed distributors. Any other use will require a separate license from a Cytiva company.
HiTrap is a registered trademark of Cytiva
MabSelect is a trademark of Cytiva
MabSelect SuRe is a trademark of Cytiva
SuRe is a trademark of Cytiva
신호어
Warning
유해 및 위험 성명서
Storage Class Code
3 - Flammable liquids
가장 최신 버전 중 하나를 선택하세요:
문서
This page shows how to perform sample desalting and buffer exchange for affinity chromatography of antibodies.