LC/MS/MS Analysis of Pesticides in Cannabis on the Ascentis® RP-Amide after QuEChERS Extraction and Cleanup Using Supel QuE Verde

application for LC-MS, application for SPE

ログイン組織・契約価格を表示する


About This Item

テクニック

LC/MS: suitable
solid phase extraction (SPE): suitable

試験パラメータ

sample preparation: Dispersive (QuEChERS)
sample/matrix: 1.9 g of cannabis spiked with pesticides at 50 ng/g
extraction tube: Supel QuE Citrate extraction tube (55227-U)
extraction process: add 10 mL of acetonitrile; shake for 30 minutes at 2500 rpm on shaker; add contents of Supel QuE citrate extraction tube; (shake immediately for 1 minute; centrifuge at 5000 rpm for 5 minutes)
clean-up tube: Supel QuE Verde, 2 mL (55227-U)
clean-up process: transfer 1 mL of acetonitrile layer into Supel QuE Verde cleanup tube; shake for 1 minutes; (centrifuge at 5000 rpm for 3 minutes; draw of supernatant for LC/MS/MS analysis)
column: Ascentis RP-Amide, 10 cm x 2.1 mm I.D., 3.0 μm particles (565301-U)
column: Ascentis RP-Amide Supelguard guard cartridge, 2 cm x 2.1 mm I.D., 5 μm particles (565372-U)
mobile phase: [A] 5 mM ammonium formate, 0.1% formic acid in 95:5 water:acetonitrile; [B] 5 mM ammonium formate, 0.1% formic acid in 5:95 water:acetonitrile
gradient: 10% B held for 1 min; to 100% B in 13 min; held at 100% B for 6 min; to 10% in 0.5 min; held at 10% B for 6 min
flow rate: 0.4 mL/min
column temp.: 30 °C
detector: MRM
injection: 5 μL
sample: QuEChERS extract of cannabis in acetonitrile

適合性

application for LC-MS
application for SPE

アプリケーション

environmental

アナリシスノート

Consumption of cannabis and/or cannabis based products is currently legal in some form in 28 US states plus the District of Columbia. Pesticide testing on plant material and extracts is required in some states, including Oregon, which mandates a specific target list for testing. A common approach used by testing labs for pesticides is QuEChERS extraction followed by a dispersive cleanup. In this application a majority of the LC-amenable pesticides on the Oregon list were extracted from cannabis spiked at 50 ng/g using QuEChERS, followed by cleanup with Supel QuE Verde. Analysis was achieved using the Ascentis® RP-Amide column equipped with a guard. Supel QuE Verde offered sufficient background reduction for LC/MS/MS analysis with recoveries of 70-120% for a majority of the pesticides.

その他情報

Sample pretreatment: Added 10 mL of deionized water, mixed and equilibrated for 10 minutes

法的情報

Ascentis is a registered trademark of Merck KGaA, Darmstadt, Germany
Supel is a trademark of Sigma-Aldrich Co. LLC

アナライト

詳細

    Daminozide
    Acephate
    Oxamyl
    Methomyl
    Thiamethoxam
    Flonicamid
    Imidacloprid
    Dimethoate
    Acetamiprid
    Aldicarb
    Thiacloprid
    Imazalil (Enilconazole)
    Dichlorvos
    Propoxur
    Spiroxamine
    Carbofuran
    Metalaxyl
    Carbaryl
    Spinosyn A
    Spinosyn D
    Spirotetramat
    Paclobutrazol
    Chlorantraniliprole
    Methiocarb (Mercaptodimethur)
    Myclobutanil
    Azoxystrobin
    Ethoprop (Ethoprophos)
    Phosmet (Imidan)
    Tebuconazole
    Boscalid (Nicobifen)
    Bifenazate (D 2341)
    Malathion
    Propiconazole
    Fenoxycarb
    Kresoxim methyl
    Clofentezin
    Prallethrin
    Trifloxystrobin
    Piperonyl butoxide
    Fenpyroximate (E)
    Hexylthiazox
    Etoxazole
    Spiromesifen
    Pyrethrin
    Pyridaben
    すべて表示 (45)

最新バージョンのいずれかを選択してください:

試験成績書(COA)

Lot/Batch Number

It looks like we've run into a problem, but you can still download Certificates of Analysis from our 資料 section.

サポートが必要な場合は、お問い合わせください カスタマーサポート

以前この製品を購入いただいたことがある場合

文書ライブラリで、最近購入した製品の文書を検索できます。

文書ライブラリにアクセスする

ライフサイエンス、有機合成、材料科学、クロマトグラフィー、分析など、あらゆる分野の研究に経験のあるメンバーがおります。.

製品に関するお問い合わせはこちら(テクニカルサービス)