KHGEA2ZTH1
ミリポアエクスプレス(Millipore Express)®SHC、オプティキャップ(Opticap)®XLTカプセル
inlet connection diam. 1.5 in., Opticap® XLT 20, pore size 0.5/0.2 μm, cartridge nominal length 20 in. (50 cm)
別名:
Autoclavable Opticap XLT 20 Millipore Express SHC 0.5/0.2 μm TC/HB 1pk w/gauge port
About This Item
おすすめの製品
物質
polyethersulfone
polypropylene
polypropylene housing
polypropylene support
polypropylene vent cap
polysulfone
silicone seal
品質水準
法規制遵守
meets FDA Indirect Food Additive requirements cited in 21 CFR 177-182 (all component materials)
無菌性
non-sterile
滅菌適合性
autoclavable compatible
製品種目
Opticap® XLT 20
特徴
hydrophilic
メーカー/製品名
Millipore Express®
パラメーター
1.0 bar max. inlet pressure (15 psi) at 80 °C
1000 bar max. differential pressure (15 psi) at 80 °C (Forward)
2.75 bar max. inlet pressure (40 psi) at 60 °C
2100 bar max. differential pressure (30 psi) at 25 °C (Reverse; intermittent)
25 °C max. inlet temp.
5.5 bar max. inlet pressure (80 psi) at 25 °C
5500 mbar max. differential pressure (80 psi) at 25 °C (Forward)
6.9 bar max. inlet pressure (100 psi) at 25 °C (intermittent)
6900 mbar max. differential pressure (100 psi) at 25 °C (Forward; intermittent)
80 psig max. inlet pressure
テクニック
sterile filtration: suitable
L
61.1 cm (24.0 in.)
W
7.8 in.
カートリッジの長さ(公称)
20 in. (50 cm)
直径
19.8 cm (7.8 in.)
ろ過面積
0.98 m2
入口接続部直径
1.5 in.
出口接続部直径
5/8 in.
不純物
≤0.25 EU/mL bacterial endotoxins (LAL test, Aqueous extraction)
<0.25 EU/mL USP bacterial endotoxins (LAL test, sample aqueous extraction)
マトリックス
Millipore Express® SHC
pore size
0.5/0.2 μm pore size
泡立ち点
≥4000 mbar (58 psig), air with water at 23 °C
フィッティング
3/4 in. gauge port (T-Line Side)
inlet sanitary flange
outlet hose barb
1/4 in. vent hose barb (with double O-ring Seal)
(38 mm (1 1/2 in.) Sanitary Flange Inlet and 16 mm (5?8 in.) Hose Barb Outlet)
詳細
包装
その他情報
- Organism Retention: Microorganism
- Mode of Action: Filtration (size exclusion)
- Application: BioProcessing
- Intended Use: Reduction or removal of microorganism/bioburden
- Instructions for Use: Please refer installation section of OPTICAP® XL Capsules and OPTICAP® XLT Capsules user guide
- Storage Statement: Store in dry location
- Disposal Statement: Dispose of in accordance with applicable federal, state and local regulations.
調製ノート
3 autoclave cycles of 60 min @ 126 °C; not in-line steam sterilizable
アナリシスノート
Samples were quantitatively retentive of a minimum Brevundimonas diminuta challenge concentration of 1 x 10⁷ CFU/cm² using ASTM® F838 methodology.
After sterilization and a controlled water flush of 40 L, samples exhibited less than 500 ppb TOC per USP <64> and less than 1.3 µS/cm per USP <64> at 25 °C.
法的情報
免責事項
Not finding the right product?
Try our 製品選択ツール.
試験成績書(COA)
製品のロット番号・バッチ番号を入力して、試験成績書(COA) を検索できます。ロット番号・バッチ番号は、製品ラベルに「Lot」または「Batch」に続いて記載されています。
ライフサイエンス、有機合成、材料科学、クロマトグラフィー、分析など、あらゆる分野の研究に経験のあるメンバーがおります。.
製品に関するお問い合わせはこちら(テクニカルサービス)