おすすめの製品
物質
aluminum hardware
認証
ISO 14698, ATEX
特徴
portable
メーカー/製品名
MAS-100 NT® Ex
アプリケーション
air monitoring
pharmaceutical
適合性
for use with 90 mm Petri dishes
詳細
The MAS-100 NT® Ex with filter is specially designed for use in explosion proof areas. The MAS-100 NT® Ex has received ATEX Conformityand can be used in zone 2 and gas groups 11A,11B, and 11C in temperature classes T1 to T4. The MAS-100 NT® Ex with filter microbial air monitoring systems are high performance, portable, battery operated instrument which are based on the principle of the Andersen air sampler. These compact yet sophisticated devices are the preferred choice for those demanding the highest quality in microbial air monitoring. A preselected amount of air is aspirated through a 300-hole perforated lid and impacted onto the surface of growth media in standard 90-100 mm Petri dishes or 55-60 mm contact plates. Microorganisms are collected on the culture media and, after an appropriate incubation period, the colonies are counted. The system measures the airflow and regulates it to a constant value of 100 standard litres/min A Standard Liter per Minute (SLPM) is adjusted to a specific environmental temperature and pressure, this ensure the amount of air sampled is always the same indipendant of environmental conditions.. These systems have the highest airflow accuracy of ± 2.5 %, compared to others that can be as high as ±10 %. The integrated flow sensor allows the user to freely interchange the perforated lids without affecting the accuracy or the calibration of the unit. The MAS-100 NT® Ex is powered by a Lithium ion rechargeable battery with an intelligent charging program that assures long battery life without routine discharging. If instruments are used in different cleanroom classes, there is a minor risk that particles may be carried over inside an instrument. In this case, a MAS-100 NT® Ex with Filter equipped with with a HEPA H13 particle filter at the exhaust prevents any potential transfer of particles from higher to lower contaminated areas. The filtration efficiency of 99.95% has been verified under worst case conditions. Furthermore the filter has no significant influence on the airflow calibration and therefore microbial sampling efficiency. In addition, stressing the filter for a simulated year of heavy usage in an uncontrolled environment does not reduce the filtration efficiency and won′t clog the filter. It is recommended to replace the filter on a yearly basis by a new HEPA H13 Filter, 74 mm for MAS-100 NT®/MAS-100 NT® Ex with filter.
アプリケーション
The MAS-100 NT® Ex w. filter microbial air sampler is used for environmental monitoring of ambient air in explosion hazard areas. Microbial air sampling devices are extensively used in Pharmaceutical and Personal care industries as well as Food, Beverage and Environmental applications.
構成
1 unit in hard carrying case, battery, 300-hole perforated lid, dust cover, mains charger, USB cable, 3 mm allen key, and USB stick with all documentation
法的情報
MAS-100 NT is a registered trademark of MBV AG
付属品
製品番号
詳細
価格
同等製品
製品番号
詳細
価格
大気捕集デバイス
試験成績書(COA)
製品のロット番号・バッチ番号を入力して、試験成績書(COA) を検索できます。ロット番号・バッチ番号は、製品ラベルに「Lot」または「Batch」に続いて記載されています。
資料
Viable Air Samplers are used for quantitative determination of airborne contamination in cleanrooms and isolators. To maintain the cleanliness of the environment, the air flow should neither disturb the unidirectional air flow nor pose a contamination risk to the environment.
ライフサイエンス、有機合成、材料科学、クロマトグラフィー、分析など、あらゆる分野の研究に経験のあるメンバーがおります。.
製品に関するお問い合わせはこちら(テクニカルサービス)