G003781
Volatile Fraction of Limoncello Using SPME on the SLB-5ms
application for SPME, suitable for GC
About This Item
tecniche
gas chromatography (GC): suitable
solid phase microextraction (SPME): suitable
Parametri analitici
sample/matrix: Limoncello diluted 1:100 in water, then 5 mL of this solution transferred to a 10 mL crimped vial spiked with 100 μL of a 100 ppm solution of nonane (internal standard)
SPME fiber: 100 μm polydimethylsiloxane (57300-U)
extraction: headspace, with a pre-equilibration period of 10 min. at 40 °C, and then a fiber exposure of 30 min. at 40 °C, agitated in an alternate clockwise-anticlockwise rotation mode at 500 ppm
desorption process: 250 °C for 5 min
column: SLB-5ms, 30 m x 0.25 mm I.D., 0.25 μm (28471-U)
oven: 40 °C, 3 °C/min to 250 °C (2 min)
inj. temp.: 250 °C
detector: FID, 260 °C (sampling rate 80 ms)
carrier gas: helium, 35.0 cm/sec
injection: splitless
liner: 0.75 mm I.D. SPME (2633901)
sample: Limoncello liquor
Compatibilità
application for SPME
applicazioni
food and beverages
Risultati analitici
Note legali
Analita
Colonna analitica
Fibra per SPME
Prodotti correlati
Standard
Scegli una delle versioni più recenti:
Certificati d'analisi (COA)
Ci dispiace, ma al momento non ci sono COA disponibili online per questo prodotto.
Se ti serve aiuto, non esitare a contattarci Servizio Clienti
Possiedi già questo prodotto?
I documenti relativi ai prodotti acquistati recentemente sono disponibili nell’Archivio dei documenti.
Il team dei nostri ricercatori vanta grande esperienza in tutte le aree della ricerca quali Life Science, scienza dei materiali, sintesi chimica, cromatografia, discipline analitiche, ecc..
Contatta l'Assistenza Tecnica.