Passa al contenuto
Merck

HPLC Analysis of Pipecolic Acid Enantiomers on Astec® CHIROBIOTIC® T

HPLC Analysis of Pipecolic Acid Enantiomers on Astec® CHIROBIOTIC® T application for HPLC

Materiali

Colonna analitica

N° Catalogo
Descrizione
Determinazione del prezzo

Colonna HPLC chirale Astec® CHIROBIOTIC® T

5 μm particle size, L × I.D. 25 cm × 4.6 mm

Standard

N° Catalogo
Descrizione
Determinazione del prezzo

Pipecolinic acid

98%

CONDITIONS

column

CHIROBIOTIC T, 25 cm x 4.6 mm I.D., 5 μm particles (12024AST)

mobile phase

[A] water; [B] acetonitrile; (50:50, A:B)

flow rate

1 mL/min

column temp.

25 °C

detector

UV, 220 nm

injection

1 μL

sample

5 mg/mL in methanol

Descrizione

Note legali

Astec is a registered trademark of Merck KGaA, Darmstadt, Germany
CHIROBIOTIC is a registered trademark of Sigma-Aldrich Co. LLC