Pular para o conteúdo
Merck

HPLC Analysis of Aminopterin and Related Compounds on Ascentis® Express C18

HPLC Analysis of Aminopterin and Related Compounds on Ascentis® Express C18 application for HPLC

Materiais

coluna analítica

Nº do produto
Descrição
Preços

Coluna de HPLC Ascentis® Express C18, 2,7 μm

2.7 μm particle size, L × I.D. 15 cm × 4.6 mm

padrão

Nº do produto
Descrição
Preços

Ácido fólico

≥97%

Metotrexato

meets USP testing specifications

CONDITIONS

column

Ascentis Express C18, 15 cm x 4.6 mm I.D., 2.7 μm particles (53829-U)

mobile phase

[A] 10 mM ammonium formate, pH 3.0; [B] acetonitrile; (90:10, A:B)

flow rate

1.0 mL/min

column temp.

35 °C

detector

UV, 280 nm

injection

5 μL

Descrição

Nota de análise

This application demonstrates the suitability of the Ascentis C18 for the analysis of aminopterin, folic acid and methotrexate. Structures along with the optimized chromatogram obtained on the Ascentis Express C18 are presented

Informações legais

Ascentis is a registered trademark of Merck KGaA, Darmstadt, Germany