794-0717
GC Analysis of Organophosphorus Pesticides (OP-Pest) on a 5% Phenyl Column after SPME using 100 μm PDMS Fiber
suitable for GC, application for SPME
About This Item
técnica(s)
gas chromatography (GC): suitable
solid phase microextraction (SPME): suitable
parâmetros de teste
sample/matrix: 50 ppb each in 1.5 mL saturated salt water, pH 7.2
SPME fiber: polydimethylsiloxane, 100 μm (57300-U)
extraction: 20 min immersion with rapid stirring
column: 5% phenyl polysiloxane, 30 m × 0.25 mm I.D., 0.25 μm (Supelco equivalent, SLB-5ms, 28471-U)
oven: 60 °C (1 min) to 300 °C at 12 °C/min, hold 5 min
inj. temp.: splitless (0.75 mm I.D. liner), 270 °C
detector: MS (m/z = 45-400, 0.6sec/scan)
carrier gas: helium, 30 cm/sec
adequação
application for SPME
aplicação(ões)
environmental
food and beverages
Analito
Fibra para SPME
coluna analítica
Escolha uma das versões mais recentes:
Certificados de análise (COA)
It looks like we've run into a problem, but you can still download Certificates of Analysis from our Documentos section.
Se precisar de ajuda, entre em contato Atendimento ao cliente
Já possui este produto?
Encontre a documentação dos produtos que você adquiriu recentemente na biblioteca de documentos.
Nossa equipe de cientistas tem experiência em todas as áreas de pesquisa, incluindo Life Sciences, ciência de materiais, síntese química, cromatografia, química analítica e muitas outras.
Entre em contato com a assistência técnica