G003784
Volatile Fraction of Roasted Coffee Beans Using SPME on Omegawax 250
application for SPME, suitable for GC
About This Item
technique(s)
gas chromatography (GC): suitable
solid phase microextraction (SPME): suitable
test parameters
sample/matrix: 2 g of coffee beans
SPME fiber: 50/30 μm divinylbenzene/Carboxen on polydimethylsiloxane on a StableFlex fiber (57329-U)
extraction: headspace, with a pre-equilibration period of 10 min. at 60 °C, and then a fiber exposure of 40 min. at 60 °C, agitated in an alternate clockwise-anticlockwise rotation mode at 500 rpm
desorption process: 260 °C for 5 min
column: Omegawax 250, 30 m x 0.25 mm I.D., 0.25 μm (24136)
oven: 40 °C (5 min), 4 °C/min to 230 °C, 50 °C/min to 280 °C (2 min)
inj. temp.: 260 °C
detector: MS, m/z 40-400
carrier gas: helium, 34.8 cm/sec constant
injection: splitless
scan range: m/z 40-400
sample: Roasted coffee
liner: 0.75 mm I.D. SPME/Splitless (2633901)
suitability
application for SPME
application(s)
food and beverages
Analysis Note
Legal Information
Analito
Related product
SPME fiber
analytical column
mobile phase component
standard
Elija entre una de las versiones más recientes:
Certificados de análisis (COA)
Lo sentimos, en este momento no disponemos de COAs para este producto en línea.
Si necesita más asistencia, póngase en contacto con Atención al cliente
¿Ya tiene este producto?
Encuentre la documentación para los productos que ha comprado recientemente en la Biblioteca de documentos.
Nuestro equipo de científicos tiene experiencia en todas las áreas de investigación: Ciencias de la vida, Ciencia de los materiales, Síntesis química, Cromatografía, Analítica y muchas otras.
Póngase en contacto con el Servicio técnico