- Effect of nitroblue tetrazolium concentration on the fructosamine assay for quantifying glycated protein.
Effect of nitroblue tetrazolium concentration on the fructosamine assay for quantifying glycated protein.
Experiments with nitroblue tetrazolium (NBT) at 0.25, 0.50, and 0.75 mmol/L confirmed that reactivity of 1-deoxy-1-morpholino-D-fructose, the standard, was more sensitive to reagent concentration than was protein ketoamine and demonstrated why an increase in NBT or a decrease in sample volume yields lower values for serum fructosamine. Analyses of clinical samples from diabetic and nondiabetic patients and from healthy subjects indicated that, with NBT at 0.25 mmol/L, discrimination between diabetics and other subjects was improved by changing the sample:reagent volume ratio from the usual 1:10 to 1:25 (9- to 12-min reading time). Discrimination improved further with NBT at 0.75 mmol/L and a 1:25 ratio but a later reading time (18-21 min); these conditions may minimize the effects of some but not all interferents.