LC/MS (TOF) Analysis of Illicit Bath Salts in Urine on Ascentis® Express HILIC after Solid Phase Extraction (SPE) using Supel™-Select SCX
application for HPLC
About This Item
metody
HPLC: suitable
parametry testowe
sample/matrix: urine samples spiked at 100 ng/mL with each target analyte (To ensure full ionization of the analytes, spiked samples were treated with formic acid to a final concentration of 0.1% formic acid.)
SPE tube/cartridge: Supel-Select SCX, 30 mg/1 mL (54240-U)
condition: 1 mL 1% formic acid acetonitrile then 1 mL water
sample addition: 1 mL spiked water blank or urine
washing: 1 mL water, 1 mL 1% formic acid acetonitrile, 1 mL water
elution: 2 mL 10% ammonium hydroxide in acetonitrile
eluate post-treatment: thoroughly mix via vortex agitation, evaporate 1 mL aliquot to dryness, reconstitute in 100 μL water:methanol
column: Ascentis Express HILIC, 10 cm x 2.1 mm I.D., 2.7 μm particles (53939-U)
mobile phase: (A) 5 mM ammonium formate acetonitrile; (B) 5 mM ammonium formate water; (98:2, A:B); premixed
flow rate: 0.6 mL/min
pressure: 1842 psi (127 bar)
column temp.: 35 °C
detector: MS, ESI(+), TIC, m/z 100-1000
injection: 1 μL
sample: 200 μg/L ea. in acetonitrile
przydatność
application for HPLC
Zastosowanie
forensics and toxicology
Komentarz do analizy
Informacje prawne
Analit
Standardowe
kolumna analityczna
probówka lub płytka SPE
składnik fazy ruchomej
Wybierz jedną z najnowszych wersji:
Certyfikaty analizy (CoA)
Przepraszamy, ale COA dla tego produktu nie jest aktualnie dostępny online.
Proszę o kontakt, jeśli potrzebna jest pomoc Obsługa Klienta
Masz już ten produkt?
Dokumenty związane z niedawno zakupionymi produktami zostały zamieszczone w Bibliotece dokumentów.
Nasz zespół naukowców ma doświadczenie we wszystkich obszarach badań, w tym w naukach przyrodniczych, materiałoznawstwie, syntezie chemicznej, chromatografii, analityce i wielu innych dziedzinach.
Skontaktuj się z zespołem ds. pomocy technicznej