G003762
HPLC Analysis of Piroxicam and 2-Aminopyridine in Urine on Discovery® HS F5 after SPE using Discovery® DSC-MCAX
application for SPE, application for HPLC
About This Item
metody
HPLC: suitable
solid phase extraction (SPE): suitable
parametry testowe
sample preparation: SPE (Solid Phase Extraction)
sample/matrix: human urine spiked with piroxicam (4 μg/mL) and 2-aminopyridine (10 μg/mL)
SPE tube/cartridge: Discovery DSC-MCAX, 100 mg/3 mL (52783-U)
condition: 1 mL methanol; 1 mL 10 mM potassium phosphate, pH 3
sample addition: 1 mL
washing: 1 mL 10 mM potassium phosphate, pH 3; 1 mL methanol
elution: 1 mL 5% ammonium hydroxide in methanol
eluate post-treatment: evaporate to dryness with nitrogen at room temperature, reconstitute in 1 mL mobile phase
column: Discovery HS F5, 15 cm x 4.6 mm I.D., 5 μm particles (567516-U)
mobile phase: [A] 10 mM potassium phosphate, pH 6: [B] acetonitrile (85:15, A:B)
flow rate: 2 mL/min
column temp.: 25 °C
detector: UV, 220 nm
injection: 10 μL
przydatność
application for HPLC
application for SPE
Zastosowanie
clinical
Zastosowanie
Piroxicam and 2-aminopyridine (piroxicam′s polar metabolite) was spiked into human urine at the levels of 4 and 10 μg/mL, respectively. The urine sample was diluted 1:1 with 10 mM potassium phosphate, pH 3.
Informacje prawne
kolumna analityczna
probówka lub płytka SPE
Wybierz jedną z najnowszych wersji:
Certyfikaty analizy (CoA)
Przepraszamy, ale COA dla tego produktu nie jest aktualnie dostępny online.
Proszę o kontakt, jeśli potrzebna jest pomoc Obsługa Klienta
Masz już ten produkt?
Dokumenty związane z niedawno zakupionymi produktami zostały zamieszczone w Bibliotece dokumentów.
Nasz zespół naukowców ma doświadczenie we wszystkich obszarach badań, w tym w naukach przyrodniczych, materiałoznawstwie, syntezie chemicznej, chromatografii, analityce i wielu innych dziedzinach.
Skontaktuj się z zespołem ds. pomocy technicznej