Przejdź do zawartości
Merck

CMP871TS1

Millipore

Milligard® PES 1.2/0.8 µm, Cartridge

pore size 1.2/0.8 μm, cartridge nominal length 10 in. (25 cm)

Synonim(y):

Milligard PES 10" 1.2/0.8 μm Cartridge

Zaloguj sięWyświetlanie cen organizacyjnych i kontraktowych


About This Item

Kod UNSPSC:
23151806
eCl@ss:
32031690

Materiały

polyethersulfone membrane
polypropylene housing
polypropylene support
polypropylene vent cap
silicone seal

Poziom jakości

zgodność regionalna

meets FDA Indirect Food Additive requirements cited in 21 CFR 177-182 (all component materials)

sterylność

non-sterile

Kompatybilność ze sterylizacją

autoclavable compatible
steam-in-place compatible

linia produktu

EMPROVE® Filter
Milligard®

Właściwości

hydrophilic

producent / nazwa handlowa

Milligard®

Parametry

1.38 bar max. differential pressure (20 psid) at 25 °C (Reverse)
5.52 bar max. differential pressure (80 psid) at 4-25 °C (Forward)

metody

prefiltration: suitable

Długość

317.5 mm (12.5 in.)

Szerokość

2.7 in.

długość nominalna kartridża

10 in. (25 cm)

Średnica

69 mm (2.71 in.)

powierzchnia filtracyjna

0.53 m2

zanieczyszczenia

<0.25 EU/mL bacterial endotoxins (LAL test, per device; aqueous extraction; meeting requirements of USP <85>, EP 2.6.14 and JP 4.01)

Matryca

Milligard® PES

pore size

1.2/0.8 μm pore size

moc wejściowa

sample type liquid

kod wkładu

Code 7

Opis ogólny

Device Configuration: Cartridge

Inne uwagi

The product is manufactured in a facility which adheres to Good Manufacturing Practices. The manufacturing facility’s Quality Management System is approved by an accredited registering body to the appropriate ISO 9001 Quality Standard.
Directions for Use

  • Mode of Action: Filtration (size exclusion)
  • Application: BioProcessing
  • Intended Use: Reduction or removal of microorganism/bioburden
  • Instructions for Use: Please refer wetting instructions of Filters with MILLIGARD® PES Membrane user guide
  • Storage Statement: Please refer Filters with MILLIGARD® PES Membrane user guide
  • Disposal Statement: Dispose of in accordance with applicable federal, state and local regulations.

Uwaga dotycząca przygotowania

Sterilization Method
Autoclave wet 25x, 60 minute cycles at 128 °C. In-line steam (wet): 10 forward cycles, 30 min., 135 °C at ≤0.34 bar (5 psid) and 5 reverse cycles, 30 min., 135 °C at ≤0.07 bar (<1 psid)*.

*Devices were validated using steam at full differential pressure starting at 31 psid to a final differential pressure of 1 psid, resulting in a calculated average of 4 psid in forward and reverse directions.
This product was manufactured with materials that meet the criteria for a "non-fiber releasing" filter as defined in 21 CFR 210.3 (b) (6).

Informacje prawne

Emprove is a registered trademark of Merck KGaA, Darmstadt, Germany
MILLIGARD is a registered trademark of Merck KGaA, Darmstadt, Germany
Ta strona może zawierać tekst przetłumaczony maszynowo.

Certyfikaty analizy (CoA)

Poszukaj Certyfikaty analizy (CoA), wpisując numer partii/serii produktów. Numery serii i partii można znaleźć na etykiecie produktu po słowach „seria” lub „partia”.

Masz już ten produkt?

Dokumenty związane z niedawno zakupionymi produktami zostały zamieszczone w Bibliotece dokumentów.

Odwiedź Bibliotekę dokumentów

Nasz zespół naukowców ma doświadczenie we wszystkich obszarach badań, w tym w naukach przyrodniczych, materiałoznawstwie, syntezie chemicznej, chromatografii, analityce i wielu innych dziedzinach.

Skontaktuj się z zespołem ds. pomocy technicznej