Sterikon® Plus Bioindicator
ampule of 2 mL, To monitor the effectiveness of the autoclaving by steam sterilization
Zaloguj sięWyświetlanie cen organizacyjnych i kontraktowych
About This Item
Polecane produkty
agency
EP
USP
Poziom jakości
Postać
liquid
Właściwości
ready-to-use
opakowanie
ampule of 2 mL
Zastosowanie
microbiology
temp. przechowywania
2-8°C
przydatność
Bacillus spp.
Opis ogólny
The Sterikon® plus Bioindicator contains a nutrient broth, sugar, a pH indicator, and spores of a non-pathogenic organism, Geobacillus stearothermophilus ATCC 7953 (sporulation optimized) in a single ampoule. The spores are completely killed upon heating in compressed steam at a temperature of 121 ± 0.5°C (245 kPa), after 15 minutes. At lower temperatures or lower exposure times, a small number of spores can survive at and can grow upon incubation. Sterikon® ampoules are placed into the autoclave along with the batch to be autoclaved. The efficiency of the sterilization process is checked by the incubation of the ampoules. Red (or red-violet) color of the ampoule contents indicates no growth of Geobacillus stearothermophilus and adequate sterilization, whereas the yellow (or yellow-orange) color of the ampoule contents indicates growth and inadequate sterilization.
Zastosowanie
Sterikon® plus Bioindicator System is used to monitor the effectiveness of the autoclaving by steam sterilization for 15 minutes at 121°C. It is suitable for checking the sterilization efficiency of the culture media products used for testing of pharmaceuticals (especially drugs in ampoule form), food and water samples, along with canned foods and other products to be autoclaved.
Powiązanie
Replaces: 1.10274.0003
Komentarz do analizy
Appearance: opalescent and reddish violet
Number of viable spores per ampoule: 5.0E+05 - 1.0E+07
Growth (35 °C): no growth
Growth (60 °C): growth
Performance test (6 min, 121 °C): Survival time 6 min/121 °C
Performance test (15 min, 121 °C): Kill time 15 min/121 °C
D value (121 °C): 1.5 - 2.0 min
Z value: 7.0 - 10.0 °C
Identity: Geobacillus stearo-thermophilus ATCC 7953
The determination of the D-value is based on the Stumbo-Murphy-Cochran method.
The identity control is carried out via MicroSeq-System.
Organism: Geobacillus stearothermophilus ATCC 7953 (sporulation optimized).
Ampoules: 2 ml.
Storage temperature: +2° to +8°C.
Number of viable spores per ampoule: 5.0E+05 - 1.0E+07
Growth (35 °C): no growth
Growth (60 °C): growth
Performance test (6 min, 121 °C): Survival time 6 min/121 °C
Performance test (15 min, 121 °C): Kill time 15 min/121 °C
D value (121 °C): 1.5 - 2.0 min
Z value: 7.0 - 10.0 °C
Identity: Geobacillus stearo-thermophilus ATCC 7953
The determination of the D-value is based on the Stumbo-Murphy-Cochran method.
The identity control is carried out via MicroSeq-System.
Organism: Geobacillus stearothermophilus ATCC 7953 (sporulation optimized).
Ampoules: 2 ml.
Storage temperature: +2° to +8°C.
Informacje prawne
STERIKON is a registered trademark of Merck KGaA, Darmstadt, Germany
This page may contain text that has been machine translated.
Kod klasy składowania
12 - Non Combustible Liquids
Klasa zagrożenia wodnego (WGK)
WGK 1
Temperatura zapłonu (°F)
Not applicable
Temperatura zapłonu (°C)
Not applicable
Certyfikaty analizy (CoA)
Poszukaj Certyfikaty analizy (CoA), wpisując numer partii/serii produktów. Numery serii i partii można znaleźć na etykiecie produktu po słowach „seria” lub „partia”.
Masz już ten produkt?
Dokumenty związane z niedawno zakupionymi produktami zostały zamieszczone w Bibliotece dokumentów.
Klienci oglądali również te produkty
Nasz zespół naukowców ma doświadczenie we wszystkich obszarach badań, w tym w naukach przyrodniczych, materiałoznawstwie, syntezie chemicznej, chromatografii, analityce i wielu innych dziedzinach.
Skontaktuj się z zespołem ds. pomocy technicznej