gas chromatography (GC): suitable
試験パラメータ
column: SLB-5ms, 30 m x 0.25 mm I.D., 0.25 μm (24103-U)
oven: 70 °C (0.2 min), 25 °C/min to 265 °C, 5 °C/min to 315 °C (2 min)
inj. temp.: 250 °C
detector: FlD, 350 °C
carrier gas: helium, 1.3 mL/min, constant
injection: 2.0 μL pulsed splitless (30 psi until 0.2 min)
liner: 4 mm I.D. single taper
sample: 1 μg/mL PAH standard in methylene chloride
アプリケーション
environmental
アナリシスノート
US EPA Method 610 describes the analysis of polynuclear aromatic hydrocarbons (commonly referred to as PAHs or PNAs) by both HPLC and GC. The method designates the use of a packed GC column, however capillary columns will provide better resolution, baseline stability, and longer lifetime. This application illustrates the separation of the target compound list in this method on the SLB-5ms capillary GC column. A pressure pulsed injection was used to improve response of the heavier PAHs, and to reduce .the larger vapor volume resulting from a 2 μL injection.
法的情報
SLB is a registered trademark of Merck KGaA, Darmstadt, Germany
アナライト
詳細
1-Fluoronaphthalene
Naphthalene
2-Fluorobiphenyl
Acenaphthylene
Acenaphthene
Fluorene
Phenanthrene
Anthracene
Fluoranthene
Pyrene
Chrysene
Benzo[b]fluoranthene
Benzo[k]fluoranthene
Indeno[1,2,3-cd]pyrene
Benzo[ghi]perylene
すべて表示 (15)