コンテンツへスキップ
Merck
すべての画像(1)

Key Documents

G005786

LC/MS/MS Analysis of Spice Cannabinoids in Plasma on Ascentis® Express F5 after Solid Phase Extraction (SPE) using HybridSPE®-Phospholipid

application for HPLC

ログイン組織・契約価格を表示する


About This Item

テクニック

HPLC: suitable

試験パラメータ

sample/matrix: rabbit plasma, unfiltered K2-EDTA spiked with Spice cannabinoids (5 ng/mL each)
SPE well plate: HybridSPE-Phospholipid, 96-well plate, 50 mg bed wt., 2 mL well vol (575656-U)
sample addition: to each well add 100 μL plasma, followed by a 300 μL of 1% formic acid in acetonitrile, agitate on orbital shaker for 2 minutes
elution: attach collection plate and apply vacuum at 10” Hg for 4 minutes
column: Ascentis Express F5, 5 cm x 2.1 mm I.D., 2.7 μm particles (53567-U)
mobile phase: [A] 10 mM ammonium formate in water, pH 6.8 (unadjusted); [B] acetonitrile; (50:50; A:B)
flow rate: 0.3 mL/min
pressure: 1296 psi (89 bar)
column temp.: 35 °C
detector: MS, ESI(+), MRM, m/z 344/155 (JWH-073 metabolite), 385/155 (JWH-200), 336/121 (JWH-250), 328/155 (JWH-073), and 342/155 (JWH-018)
injection: 2 μL

適合性

application for HPLC

アプリケーション

forensics and toxicology

アナリシスノート

Synthetic cannabinoids (Spice) are a relatively new type of designer drug used as a pseudo-legal means to get a cannabis-type high. New synthetic cannabinoids are continually being introduced as suppliers tweak the molecular structures. The ability to rapidly and reliably identify the continually changing population of these compounds in the blood or urine suspected users is a significant analytical challenge facing forensic chemists. A four prong approach using column selectivity, high purity solvents, effective sample prep, and reference standards, was used to develop a method to rapidly isolate and identify Spice cannabinoids from plasma. The Ascentis Express F5 column gave the necessary resolution, and the highest grade LC-MS solvents and additives gave adduct-free response for maximum sensitivity. Sample prep employing the HybridSPE-Phospholipid was rapid and effective and the Cerilliant reference standards enabled confident identification.

法的情報

Ascentis is a registered trademark of Merck KGaA, Darmstadt, Germany
HybridSPE is a registered trademark of Merck KGaA, Darmstadt, Germany

アナライト

詳細

    JWH-073 N-(3-hydroxybutyl) metabolite solution, 100 μg/mL in methanol, ampule of 1 mL, certified reference material, Cerilliant®
    JWH-073 solution, 100 μg/mL in acetonitrile, ampule of 1 mL, certified reference material, Cerilliant®
    JWH-018 solution, 100 μg/mL in methanol, ampule of 1 mL, certified reference material, Cerilliant®

最新バージョンのいずれかを選択してください:

試験成績書(COA)

Lot/Batch Number

申し訳ございませんが、現在この製品のCOAをオンラインで入手できません。

サポートが必要な場合は、お問い合わせください カスタマーサポート

以前この製品を購入いただいたことがある場合

文書ライブラリで、最近購入した製品の文書を検索できます。

文書ライブラリにアクセスする

ライフサイエンス、有機合成、材料科学、クロマトグラフィー、分析など、あらゆる分野の研究に経験のあるメンバーがおります。.

製品に関するお問い合わせはこちら(テクニカルサービス)