テクニック
HPLC: suitable
solid phase extraction (SPE): suitable
試験パラメータ
sample preparation: SPE (Solid Phase Extraction)
sample/matrix: horse plasma spiked at 10 μg/mL
SPE tube/cartridge: [TM"HybridSPE"]-Phospholipid Plate, 50mg/96-Well (575656-U)
sample addition: 3:1, (1% formic acid in acetonitrile); plasma
column: Ascentis Express F5, 10 cm x 2.1 mm I.D., 2.7 μm particles (53569-U)
mobile phase: [A] acetonitrile [B] 20 mM ammonium formate; (90:10, A:B); pH 4.5 with formic acid, after mixing A & B
flow rate: 0.4 mL/min
pressure: 1305 psi (90 bar)
column temp.: 35 °C
detector: MS, ESI(+), TIC, m/z 100-1000
injection: 1 μL
適合性
application for HPLC
application for SPE
アプリケーション
forensics and toxicology
アナリシスノート
Dermorphin is a natural heptapeptide opioid. Because of poor recovery from horse plasma following protein precipitation, alternate methods of extraction are sought. While the HybridSPE protocol successfully removes proteins and phospholipids, dermorphin solubility in plasma remains a limitation. Dermorphin is effectively retained on the pentafluorophenyl phase under aqueous normal phase conditions.
法的情報
Ascentis is a registered trademark of Merck KGaA, Darmstadt, Germany
HybridSPE is a registered trademark of Merck KGaA, Darmstadt, Germany
最新バージョンのいずれかを選択してください:
ライフサイエンス、有機合成、材料科学、クロマトグラフィー、分析など、あらゆる分野の研究に経験のあるメンバーがおります。.
製品に関するお問い合わせはこちら(テクニカルサービス)