テクニック
gas chromatography (GC): suitable
試験パラメータ
column: SLB-5ms, 30 m x 0.25 mm I.D, 0.25 μm (28471-U)
oven: 100 °C (1 min), 10 °C/min to 300 °C (5 min)
inj. temp.: 250 °C
MSD interface: 300 °C
scan range: selected ion monitoring (SIM), 7 monitoring groups used
carrier gas: helium, 1 mL/min constant
injection: 1.0 μL pulsed (20 psi until 0.20 min), splitless (1.0 min)
liner: 4 mm I.D., single taper
sample: 29-component pesticide mix plus 1 internal standard, each at 100 ppb in toluene
アプリケーション
agriculture
food and beverages
アナリシスノート
This application illustrates the analysis of 29 different agricultural pesticides of different classifications and functionalities. The analysis was done by GC-MS using selected ion monitoring (SIM) to allow detection of all compounds at 100 μg/L. The
standard solution was made in toluene to increase it stability and enhance response of some of the pesticides. The extra peaks in the first portion of the run are due to interferences from the solvent.
standard solution was made in toluene to increase it stability and enhance response of some of the pesticides. The extra peaks in the first portion of the run are due to interferences from the solvent.
法的情報
SLB is a registered trademark of Merck KGaA, Darmstadt, Germany
アナライト
詳細
Dichlorvos
Acephate
Propoxur
Hexachlorobenzene
Lindane
Diazinon
Chlorothalonil
Carbaryl
Dichlofluanid
Chlorpyrifos
Cyprodinil
Tolylfluanid
Captan
Thiabendazole
Folpet
cis-Chlordane
4,4′-DDE
Imazalil
Endosulfan sulfate
Dicofol
cis-Permethrin
trans-Permethrin
Coumaphos
すべて表示 (23)
最新バージョンのいずれかを選択してください:
ライフサイエンス、有機合成、材料科学、クロマトグラフィー、分析など、あらゆる分野の研究に経験のあるメンバーがおります。.
製品に関するお問い合わせはこちら(テクニカルサービス)