GC Analysis of Agricultural Pesticides (Standard) on SLB®-5ms

suitable for GC

ログイン組織・契約価格を表示する


About This Item

テクニック

gas chromatography (GC): suitable

試験パラメータ

column: SLB-5ms, 30 m x 0.25 mm I.D, 0.25 μm (28471-U)
oven: 100 °C (1 min), 10 °C/min to 300 °C (5 min)
inj. temp.: 250 °C
MSD interface: 300 °C
scan range: selected ion monitoring (SIM), 7 monitoring groups used
carrier gas: helium, 1 mL/min constant
injection: 1.0 μL pulsed (20 psi until 0.20 min), splitless (1.0 min)
liner: 4 mm I.D., single taper
sample: 29-component pesticide mix plus 1 internal standard, each at 100 ppb in toluene

アプリケーション

agriculture
food and beverages

アナリシスノート

This application illustrates the analysis of 29 different agricultural pesticides of different classifications and functionalities. The analysis was done by GC-MS using selected ion monitoring (SIM) to allow detection of all compounds at 100 μg/L. The
standard solution was made in toluene to increase it stability and enhance response of some of the pesticides. The extra peaks in the first portion of the run are due to interferences from the solvent.

法的情報

SLB is a registered trademark of Merck KGaA, Darmstadt, Germany

アナライト

詳細

    Dichlorvos
    Acephate
    Propoxur
    Hexachlorobenzene
    Lindane
    Diazinon
    Chlorothalonil
    Carbaryl
    Dichlofluanid
    Chlorpyrifos
    Cyprodinil
    Tolylfluanid
    Captan
    Thiabendazole
    Folpet
    cis-Chlordane
    4,4′-DDE
    Imazalil
    Endosulfan sulfate
    Dicofol
    cis-Permethrin
    trans-Permethrin
    Coumaphos