G006151
UHPLC/MS Analysis of Antineoplastic and Antidepressant Drugs in Plasma on Titan C18 after SPE using HybridSPE®-PLus
application for HPLC, application for SPE
About This Item
Empfohlene Produkte
Methode(n)
HPLC: suitable
solid phase extraction (SPE): suitable
Testparameter
sample/matrix: rat plasma
SPE well plate: HybridSPE-PLus 96-Well Plate, 50 mg/2 mL per well (575659-U)
sample preparation: final analyte concentration 50 ng/mL
sample addition: 400 μL of 1% formic acid acetonitile:spiked plasma (3:1) per well
column: Titan C18, 5 cm x 2.1 mm I.D., 1.9 μm particles (577122-U)
mobile phase: [A] 5 mM ammonium formate; [B] 5 mM ammonium formate in 90:10 (v:v) acetonitrile:water
gradient: 35 to 90% B in 1.5 min, held at 90% B for 0.4 min
flow rate: 0.4 mL/min
pressure: 3582 psi (247 bar)
column temp.: 35 °C
detector: TOF/MS
injection: 1 μL
Eignung
application for HPLC
application for SPE
Anwendung(en)
pharmaceutical (small molecule)
Hinweis zur Analyse
Rechtliche Hinweise
Analyt
SPE-Röhrchen oder -Platte
Standard
Hier finden Sie alle aktuellen Versionen:
Analysenzertifikate (COA)
Leider sind derzeit keine COAs für dieses Produkt online verfügbar.
Wenn Sie Hilfe benötigen, wenden Sie sich bitte an Kundensupport
Besitzen Sie dieses Produkt bereits?
In der Dokumentenbibliothek finden Sie die Dokumentation zu den Produkten, die Sie kürzlich erworben haben.
Unser Team von Wissenschaftlern verfügt über Erfahrung in allen Forschungsbereichen einschließlich Life Science, Materialwissenschaften, chemischer Synthese, Chromatographie, Analytik und vielen mehr..
Setzen Sie sich mit dem technischen Dienst in Verbindung.