LC/MS Analysis of Erythromycin Analogs on Ascentis® Express HILIC
Materiali
Colonna analitica
N° Catalogo
Descrizione
Determinazione del prezzo
Componente della fase mobile
N° Catalogo
Descrizione
Determinazione del prezzo
CONDITIONS
column
Ascentis Express HILlC, 15 cm x 4.6 mm I.D., 2.7 μm particles (53981-U)
mobile phase
[A] 100 mM ammonium formate, pH 3.0 (titrated with formic acid); [B] acetonitrile; (10:90, A:B)
flow rate
1.0 mL/min
pressure
1247 psi (86 bar)
column temp.
35 °C
detector
ESI(+); SIR
injection
5 μL
Descrizione
Risultati analitici
This application shows the suitability of Ascentis Express Silica columns for the analysis of erythromycin analogs in aqueous normal phase/HILIC mode of chromatography. Other relatively hydrophobic aminoglycosides have also been shown to separate under these conditions.
Note legali