HPLC Analysis of Tamoxifen E/Z isomers and Related Compounds on Ascentis® Express RP-Amide
Materiali
Colonna analitica
N° Catalogo
Descrizione
Determinazione del prezzo
Colonna per HPLC Ascentis® Express RP-Amide, 2,7 μm
2.7 μm particle size, L × I.D. 10 cm × 3 mmCONDITIONS
column
Ascentis Express RP-Amide, 10 cm x 3 mm I.D., 2.7 μm particles (53918-U)
mobile phase
[A] 0.1% formic acid in water; [B] 0.1% formic acid in acetonitrile
gradient
held at 25% B for 2 min; 25 to 40% B in 15 min; 40 to 25% B in 0.1 min; held at 25% B for 2.9 min
flow rate
0.6 mL/min
pressure
2292 psi (158 bar)
column temp.
35 °C
detector
UV, 260 nm
injection
2 μL
sample
10 mg/L in 10:90, water: methanol
Descrizione
Note legali
Ascentis is a registered trademark of Merck KGaA, Darmstadt, Germany