GC Analysis of Gasoline Residue in Arson Samples on a 100% Methyl Column after SPME using 100 μm PDMS Fiber
Materiali
Colonna analitica
N° Catalogo
Descrizione
Determinazione del prezzo
Fibra per SPME
N° Catalogo
Descrizione
Determinazione del prezzo
Unità fibra SPME polidimetilsilossano (PDMS)
df 100 μm(PDMS, needle size 24 ga, for use with manual holderCONDITIONS
SPME fiber
polydimethylsiloxane phase fiber, 100 μm headspace sampling (20 min) 10 sec desorption (splitless mode) (57300-U)
column
100% methyl polysiloxane, 30 m × 0.25 mm I.D., 0.25 μm (Supelco equivalent, Equity-1, 28046-U)
oven
35 °C (2 min) to 220 °C at 10 °C/min, hold 2 min, to 300 °C at 30 °C/min, hold 5 min
inj. temp.
splitless (closed 3 min), 220 °C (2 mm I.D. injector liner)
detector
FID, 300 °C
carrier gas
helium, 1 mL/min (split 50:1)