G005682

HPLC Analysis of Caffeine and Taurine on Ascentis® Express HILIC Column using Evaporative Light Scattering Detection (ELSD)

application for HPLC

Autenticatiper visualizzare i prezzi riservati alla tua organizzazione & contrattuali


About This Item

tecniche

HPLC: suitable

Parametri analitici

column: Ascentis Express HILlC, 5 cm x 3.0 mm I.D., 2.7 μm particles (53967-U)
mobile phase: [A] 100 mM ammonium formate, pH 3.0 with formic acid; [B] acetonitrile; (10:90, A:B)
flow rate: 0.6 mL/min
column temp.: 25 °C
detector: ELSD, 55 °C, 3.5 bar nitrogen
injection: 2 μL
sample: 1000 μg/mL in 10:90, water: acetonitrile

Compatibilità

application for HPLC

applicazioni

vitamins, nutraceuticals, and natural products

Risultati analitici

This application demonstrates the suitability of Ascentis Express HILIC for the efficient separation of caffeine and taurine. Elution is by aqueous normal phase (ANP) as an increase in organic will increase retention.

Note legali

Ascentis is a registered trademark of Merck KGaA, Darmstadt, Germany

Analita

Descrizione

    Caffeine
    Taurine

Colonna analitica

N° Catalogo
Descrizione
Determinazione del prezzo

Standard

N° Catalogo
Descrizione
Determinazione del prezzo

Scegli una delle versioni più recenti:

Certificati d'analisi (COA)

Lot/Batch Number

Ci dispiace, ma al momento non ci sono COA disponibili online per questo prodotto.

Se ti serve aiuto, non esitare a contattarci Servizio Clienti

Possiedi già questo prodotto?

I documenti relativi ai prodotti acquistati recentemente sono disponibili nell’Archivio dei documenti.

Visita l’Archivio dei documenti

Il team dei nostri ricercatori vanta grande esperienza in tutte le aree della ricerca quali Life Science, scienza dei materiali, sintesi chimica, cromatografia, discipline analitiche, ecc..

Contatta l'Assistenza Tecnica.