KPHLS10FF1
Capsule Opticap XL ®, filtro Durapore® da 0,45 µm
Opticap® XL 10, cartridge nominal length 13.2 in. (33.5 cm), filter diam. 4.2 in. (10.7 cm)
About This Item
Prodotti consigliati
Materiali
polyester support
polyethylene support
polypropylene filter
polypropylene housing
silicone outer O-ring
silicone-coated EPDM rubber lead O-ring
Livello qualitativo
Conformità normativa
FDA 21CFR177-182 (all component materials)
Sterilità
sterile; irradiated
Nome Commerciale
Opticap® XL 10
Caratteristiche
hydrophilic
Produttore/marchio commerciale
Opticap®
Parametri
≤15.0 mL/min air diffusion at 3.9 bar (56 psig) and 23 °C (in water)
1 bar max. inlet pressure (15 psi) at 80 °C
2.1 bar max. differential pressure (30 psid) at 4-40 °C (Reverse:)
2.8 bar max. inlet pressure (40 psi) at 60 °C
25 °C max. inlet temp.
4.1 bar max. differential pressure (60 psid) at 25 °C (Forward:)
5.5 bar max. inlet pressure (80 psi) at 23 °C
80 psig max. inlet pressure
tecniche
bioburden reduction: suitable
Lunghezza nominale della cartuccia
13.2 in. (33.5 cm)
device L
33.5 cm (13.2 in.)
device size
10 in.
Diam. filtro
4.2 in. (10.7 cm)
filter filtration area
0.62 m2
Diam. connessione ingresso
3/4 in.
inlet to outlet width
33.5 cm (13.2 in.)
Diam. connessione uscita
3/4 in.
Impurezze
≤0.25 EU/mL bacterial endotoxins (LAL test, aqueous extraction)
<0.25 EU/mL bacterial endotoxins (LAL test, sample aqueous extraction)
Matrice
Durapore®
Dimensione pori
0.45 μm pore size
Punto di bolla
≥1930 mbar (28 psig), air with water at 23 °C
Raccordi
1/4 in. fitting (vent/drain hose barb with double O-ring seal)
(19 mm (3/4 in.) Sanitary Flange Inlet and Outlet)
Cerchi prodotti simili? Visita Guida al confronto tra prodotti
Descrizione generale
Nota sulla preparazione
Gamma compatible to 40 kGy. 3 autoclave cycles of 60min at 126°C; not in-line steam sterilizable
This product was manufactured with a Durapore® membrane which meets the criteria for a "non-fiber releasing" filter as defined in 21CFR210.3(b)(6).
Risultati analitici
Altre note
- Organism Retention: Microorganism
- Mode of Action: Filtration (size exclusion)
- Application: BioProcessing
- Intended Use: Reduction or removal of microorganism/bioburden
- Instructions for Use: Please refer installation section of Opticap® XL Capsules and Opticap® XLT Capsules user guide
- Storage Statement: Store in dry location
- Disposal Statement: Dispose of in accordance with applicable federal, state and local regulations
Note legali
Non trovi il prodotto giusto?
Prova il nostro Motore di ricerca dei prodotti.
Certificati d'analisi (COA)
Cerca il Certificati d'analisi (COA) digitando il numero di lotto/batch corrispondente. I numeri di lotto o di batch sono stampati sull'etichetta dei prodotti dopo la parola ‘Lotto’ o ‘Batch’.
Possiedi già questo prodotto?
I documenti relativi ai prodotti acquistati recentemente sono disponibili nell’Archivio dei documenti.
Il team dei nostri ricercatori vanta grande esperienza in tutte le aree della ricerca quali Life Science, scienza dei materiali, sintesi chimica, cromatografia, discipline analitiche, ecc..
Contatta l'Assistenza Tecnica.