LNEBIO301
Glucose, Creatinine, Glycerides, Cholesterol, Urea and Uric acid in human frozen serum
LNE Certified Reference Material
Faça loginpara ver os preços organizacionais e de contrato
About This Item
Produtos recomendados
Formulário
liquid
prazo de validade
limited shelf life, expiry date on the label
composição
human serum, 100%
IVD
not for in vitro diagnostic use
temperatura de armazenamento
−70°C
Procurando produtos similares? Visita Guia de comparação de produtos
Descrição geral
LNE CRM Bio 301 is a higher-order certified reference material produced by the French metrological institute LNE in compliance with ISO 17034 requirements. This secondary standard is intended for use as quality control material to assess the measurement bias or measurement uncertainty of established or new measurement procedures for the determination of creatinine, glucose, total cholesterol, total glycerides, HDL-cholesterol and LDL-cholesterol in human serum.
The Certified Reference Material LNE CRM Bio 301 corresponds to a pool of Human Frozen Serum that was prepared according to the Clinical Laboratory Standards Institute (CLSI) C37-A guidelines describing the preparation and validation of commutable frozen human serum pools as Secondary Reference Materials for Cholesterol Measurement. Samples commutability was assessed through a commutability study involving several clinical laboratories. Commutability results are reported in the certificate of analysis.
Cert_ LNE CRM HbA1c 401
The Certified Reference Material LNE CRM Bio 301 corresponds to a pool of Human Frozen Serum that was prepared according to the Clinical Laboratory Standards Institute (CLSI) C37-A guidelines describing the preparation and validation of commutable frozen human serum pools as Secondary Reference Materials for Cholesterol Measurement. Samples commutability was assessed through a commutability study involving several clinical laboratories. Commutability results are reported in the certificate of analysis.
Cert_ LNE CRM HbA1c 401
Aplicação
The LNE CRM HbA1c 401 is a reference material designed for use as quality control material. It is intended for evaluating the measurement bias or uncertainty of existing or new measurement procedures for determining HbA1c in hemolysate or whole blood.
Características e benefícios
- A reliable Certified Reference Materials (CRM) for clinically relevant biomarkers
- Concentrations of glucose, creatinine, total cholesterol, triglyceride (TG), high-density lipoprotein (HDL)-cholesterol, low-density lipoprotein (LDL)-cholesterol, urea, and uric acid, are certified with higher-order reference methods.
- Derived from real patient samples.
- This CRM is pooled, tested, and certified by the LNE, which includes extensive commutability studies.
- The accuracy of the reference methods used to certify target values is verified using certified matrix reference materials.
- This CRM was used for national quality control organized by the ANSM in 2016 to assess the accuracy of some general biochemistry tests.
Outras notas
- Details on storage, safety, usage, and operating instructions are provided in the comprehensive Certificate of Analysis.
- Concentrations of glucose, total glycerides, and total cholesterol are certified by reference measurement procedures relying on Gas Chromatography Isotope Dilution mass Spectrometry (GC-IDMS) listed in JCTLM database.
- Concentrations of HDLC and LDL-C, are certified at the CDC by Beta Quantification with cholesterol being determined by the Abell Kendall method.
- Concentrations of creatinine and uric acid are certified by reference measurement procedures relying on Liquid Chromatography Isotope Dilution mass Spectrometry (LC-IDMS) listed in JCTLM database.
- This clinical matrix CRM is for research and clinical testing applications.
Código de classe de armazenamento
10 - Combustible liquids
Classe de risco de água (WGK)
WGK 1
Ponto de fulgor (°F)
Not applicable
Ponto de fulgor (°C)
Not applicable
Escolha uma das versões mais recentes:
Certificados de análise (COA)
Não está vendo a versão correta?
Se precisar de uma versão específica, você pode procurar um certificado específico pelo número do lote ou da remessa.
Já possui este produto?
Encontre a documentação dos produtos que você adquiriu recentemente na biblioteca de documentos.
Nossa equipe de cientistas tem experiência em todas as áreas de pesquisa, incluindo Life Sciences, ciência de materiais, síntese química, cromatografia, química analítica e muitas outras.
Entre em contato com a assistência técnica