GC Analysis of Sulfur Compounds in Gasoline on SPB®-1 SULFUR

suitable for GC

Faça loginpara ver os preços organizacionais e de contrato


About This Item

técnica(s)

gas chromatography (GC): suitable

parâmetros de teste

column: SPB-1 SULFUR, 30 m x 0.32 mm I.D., 4.0 μm (24158)
oven: 35 ºC, 5 ºC/min to 200 ºC
inj. temp.: 275 °C
detector: FID, 300 ºC; FPD, 300 ºC
carrier gas: helium, 20 cm/sec
injection: 1 μL 100:1 split
sample: Gasoline

aplicação(ões)

petroleum

Nota de análise

Two issues must be overcome for the analysis of sulfur compounds in fuels. 1) Detection of these compounds in a hydrocarbon sample matrix, and 2) adsorption of these compounds by the column. Sulfur specific detectors, such as a flame photometric detector (FPD), can be used to overcome the first challenge. A low phase ratio column, such as the SPB-1 SULFUR, can be used to overcome the second challenge. This slide shows the analysis of a gasoline sample, in which the column effluent was split to two detectors. The top chromatogram shows the FID response, and the bottom chromatogram shows the FPD response.

Informações legais

SPB is a registered trademark of Merck KGaA, Darmstadt, Germany

coluna analítica


Escolha uma das versões mais recentes:

Certificados de análise (COA)

Lot/Batch Number

Lamentamos, não temos COA para este produto disponíveis online no momento.

Se precisar de ajuda, entre em contato Atendimento ao cliente

Já possui este produto?

Encontre a documentação dos produtos que você adquiriu recentemente na biblioteca de documentos.

Visite a Biblioteca de Documentos

Nossa equipe de cientistas tem experiência em todas as áreas de pesquisa, incluindo Life Sciences, ciência de materiais, síntese química, cromatografia, química analítica e muitas outras.

Entre em contato com a assistência técnica