G005748

HPLC Analysis of Tamoxifen E/Z isomers and Related Compounds on Ascentis® Express RP-Amide

application for HPLC

Faça loginpara ver os preços organizacionais e de contrato


About This Item

técnica(s)

HPLC: suitable

parâmetros de teste

column: Ascentis Express RP-Amide, 10 cm x 3 mm I.D., 2.7 μm particles (53918-U)
mobile phase: [A] 0.1% formic acid in water; [B] 0.1% formic acid in acetonitrile
gradient: held at 25% B for 2 min; 25 to 40% B in 15 min; 40 to 25% B in 0.1 min; held at 25% B for 2.9 min
flow rate: 0.6 mL/min
pressure: 2292 psi (158 bar)
column temp.: 35 °C
detector: UV, 260 nm
injection: 2 μL
sample: 10 mg/L in 10:90, water: methanol