parâmetros de teste

sample/matrix: 4 mL (water + 100ng/mL each analyte 1g/mL NaCl), pH 2, in 4.6 mL vial
SPME fiber: polyacrylate, 85 μm (57304)
extraction: immersion, ambient temp., 50 min (constant stirring)
desorption process: 5 min, 230 °C
column: 5% phenyl polysiloxane, 30 m × 0.25 mm I.D., 0.25 μm (Supelco equivalent, SLB-5ms, 28471-U)
oven: 40 °C (5 min) to 100 °C at 30 °C/min, to 275 °C at 5 °C/min
inj. temp.: splitless, 230 °C
detector: MS (Ion Trap Manifold: 250°C; scan range m/z=45-400 at 0.6 sec/scan)
carrier gas: helium, 40 cm/sec, set at 40 °C

adequação

application for SPME

aplicação(ões)

environmental
food and beverages

Analito

Descrição

    Atrazine
    Prowl®, analytical standard
    Hexazinone
    Oxadiazon
    Simazine
    Terbacil
    Trifluralin
    Bromacil
    Propazine
    Propachlor
    Eptam®, PESTANAL®, analytical standard
    Ver tudo (11)

Escolha uma das versões mais recentes:

Certificados de análise (COA)

Lot/Batch Number

Lamentamos, não temos COA para este produto disponíveis online no momento.

Se precisar de ajuda, entre em contato Atendimento ao cliente

Já possui este produto?

Encontre a documentação dos produtos que você adquiriu recentemente na biblioteca de documentos.

Visite a Biblioteca de Documentos

Nossa equipe de cientistas tem experiência em todas as áreas de pesquisa, incluindo Life Sciences, ciência de materiais, síntese química, cromatografia, química analítica e muitas outras.

Entre em contato com a assistência técnica