794-0199
GC Analysis of Phenols in Water on a 5% Phenyl Column after SPME using 85 μm Polyacrylate Fiber
suitable for GC, application for SPME
About This Item
técnica(s)
gas chromatography (GC): suitable
solid phase microextraction (SPME): suitable
parâmetros de teste
sample/matrix: 50 ppb phenols in water, 27% NaCl, pH 2
SPME fiber: polyacrylate, 85 μm (57304)
extraction: 20 min immersed in water with rapid stirring
desorption process: splitless, 280 °C (closed 3 min)
column: 5% Phenyl Polysiloxane, 30 m × 0.25 mm I.D., 0.25 μm (Supelco equivalent, SLB-5ms, 28471-U)
oven: 40 °C (4 min), 12 °C/min to 260 °C
detector: MS, Scan Range m/z = 45-465 at 0.6 sec/scan
carrier gas: helium, 40 cm/sec at 40 °C
adequação
application for SPME
aplicação(ões)
chemicals and industrial polymers
environmental
food and beverages
Analito
produto relacionado
Escolha uma das versões mais recentes:
Certificados de análise (COA)
Lamentamos, não temos COA para este produto disponíveis online no momento.
Se precisar de ajuda, entre em contato Atendimento ao cliente
Já possui este produto?
Encontre a documentação dos produtos que você adquiriu recentemente na biblioteca de documentos.
Nossa equipe de cientistas tem experiência em todas as áreas de pesquisa, incluindo Life Sciences, ciência de materiais, síntese química, cromatografia, química analítica e muitas outras.
Entre em contato com a assistência técnica